The Elekta MRL Biomarkers working
group diffusion-weighted imaging (DWI) sequence (ss-EPI, b = 0, 150, 500 s/mm2,
TE/TR = 65/2931 ms, voxel size = 1.9 mm x 1.9 mm x 4 mm) was tested for
accuracy and repeatability over 6 monthly QA sessions using the QIBA diffusion
phantom.
The sequence was run on five
patients undergoing conventional treatment and four patients undergoing SABR
treatment recruited under the MOMENTUM study. The purpose of the DWI imaging
was explained to patients who then provided consent to undergo additional imaging
for research purposes after daily treatments. At each treatment, patients could
opt out of this imaging.
The T2w anatomical planning scans
were rigidly registered to the daily T2w images and the contours were deformed
across in the Monaco treatment planning system, and then rigidly registered and
copied to the daily DWI b0 images.
ADC maps were generated from b =
150, 500 s/mm2 images, and median values were extracted from whole
prostate ROIs, obtained by eroding registered treatment contours by 2 voxels so
surrounding tissue was not included in the ROI.
A 2-sided, paired t-test was used
to determine if there were any significant ADC changes between the start and
end of treatment for either treatment (null hypothesis: no change between start/end
of treatment, p < 0.05 required to reject the null hypothesis).