Copenhagen, Denmark
Onsite/Online

ESTRO 2022

Session Item

Dosimetry
6034
Poster (digital)
Physics
Evaluation of a new film dosimetry system for radiotherapy; orthochromic
Antony Palmer, United Kingdom
PO-1532

Abstract

Evaluation of a new film dosimetry system for radiotherapy; orthochromic
Authors:

Antony Palmer1, David Nash2

1Portsmouth Hospitals University NHS Trust, Medical Physics Dept., Portsmouth, United Kingdom; 2Portsmouth Hospitals University NHS Trust, Medical Physics Dept, Portsmouth, United Kingdom

Show Affiliations
Purpose or Objective

Film dosimetry is an important technique for the verification of complex radiotherapy delivery for audit or individual patient plan quality assessment. However, uncertainties including lateral scan artefact inherently limit achievable accuracy with conventional transparent radiochromic films. This study is the first to consider a completely new film and associated analysis software: Orthochromic (OrthoChrome Inc., USA), which mitigates many of these limitations. We characterise the film response to radiotherapy irradiation, evaluate improvements in the scanning process, and investigate results with a test treatment plan verification. 

Material and Methods

Orthochromic film OC-1 and Orthochrome Pro software (OrthoChrome Inc.) were evaluated in this study. The opaque film is scanned in reflective mode on a conventional flatbed scanner (Epson 12000XL). We characterised the film for use in routine radiotherapy clinic dosimetry applications and plan verifications: Including calibration accuracy, response as a function of dose level (0 to 200 Gy) and dose rate (0.4 to 2.4 Gy/min), effects of film orientation on scanner, lateral position of film on scanner, with repeated scanning, and post-exposure changes with time. The film was used in an IMRT Thorax Phantom (CIRS, USA) to verify the treatment delivery of a 10 MV FFF (flattening filter free) Pinnacle (Philips, NL) lung SABR plan (54Gy/3 fractions), with the film positioned in a transverse plane through the target. Profile comparison and gamma analysis within specified isodose (50% and 5%) were used.

Results

A reliable calibration of the OC-1 film was obtained using the Orthochrome Pro software: individual calibrations were obtained for multiple lateral scan positions on the scanner. The film exhibited a logarithmic post exposure darkening, and non-linear response as a function of dose level, consistent with radiochromic film. There were no dose rate effects observed (<0.1% change). There were no film orientation effects at scanning on pixel value (<0.1% change), and much reduced lateral scan effects across the scanner compared to radiochromic film (<0.7% change across scanner at up to 20 Gy). Repeated scanning of the film up to the tested 20 times did not change the scanned pixel value. Comparison of the measured and planned dose distributions for the SABR plan showed good agreement. Comparison profiles through the high dose region are shown in Figure 1. The gamma comparison between measured and planned is shown in figure 2.



Conclusion

We are the first to report a full characterisation of the new orthochromic film dosimetry system for radiotherapy, including an example treatment plan verification. The data provided is valuable for implementation of the system in a radiotherapy clinic or for dosimetric audits.