Thirty-three CT were analyzed (breast protocol n=25, thoracic protocol n=8). DSC and DT are showed in Fig. 1 and 2.
Minimal differences were showed if we consider DSC of lungs (0.97 for both left and right lung and DT, respectively, 17.91 and 19.13 mm) and heart (mean DSC and 95% DT were 0.93 and 10.05 mm)
The comparison of spinal cord delineation showed slightly worse DSC (0.85) and DT 6.4 mm), probably related to better accuracy of Artificial Intelligence in intercept density differences compared to human eye.
Thyroid (n=25) was the organ with more noticeable differences: in this case the DSC and DT were respectively 0.7 and 9.87 mm.
Esophagus, trachea and aorta were contoured on 8 scans. Their DSC were respectively 0.77, 0.82, 0.84, and DT were 14.81, 12.55 and 95.30 mm.
No unacceptable AC were noticed.