Copenhagen, Denmark
Onsite/Online

ESTRO 2022

Session Item

Monday
May 09
09:00 - 10:00
Poster Station 1
17: Treatment planning
Christoph Schneider, The Netherlands
3160
Poster Discussion
Physics
Artificial Intelligence and Multi-Criteria Optimization for automatic treatment plan generation
Gwenaelle Sidorski, France
PD-0734

Abstract

Artificial Intelligence and Multi-Criteria Optimization for automatic treatment plan generation
Authors:

Jocelyne MAZURIER1, Gwenaelle SIDORSKI1, Xavier Franceries2, Isabelle BERRY3, Baptiste PICHON1, Baptiste PINEL1, Igor LATORZEFF1, Olivier Gallocher4, Gaelle JIMENEZ1, Jeremy CAMILLERI1, Vincent Connord1, Yoan MARTY1, Nicolas MATHY1, Daniel ZARATE1

1Clinique Pasteur, radiotherapy, Toulouse, France; 2Université Toulouse Sabatier 3, INSERM UMR1037, TOULOUSE, France; 3CNRS, CERCO, TOULOUSE, France; 4Clinique Pasteur, radiotherapy, toulouse, France

Show Affiliations
Purpose or Objective

In RayStation v11A® TPS, 3 methods are available to generate treatment plans: A Multi-Criteria Optimization (MCO) and two Artificial Intelligences (AI), a Machine Learning (ML) and a Deep Learning (DL) algorithm. The goal is to evaluate these advanced methods to automatize our clinical routine.

Material and Methods

The ML is based on Atlas of Random Forest and the DL on a Fully Convolutional Neural Network. They generate a predicted dose distribution. The AI methods were both trained on local treatment plans database, then the obtained models were adjusted in a dedicated RaySearch module, RayLearner.

The MCO algorithm generate treatment plans (from users constraints) were no optimization goals can be improved without deteriorating another: Optimal Pareto Plans. The chosen predicted dose distribution calculated from theorical fluence.

Then, the three methods use the same Mimicking algorithm which transforms the predicted dose distribution into a deliverable treatment plan.

Methods have been evaluated on prostate cancer treated at 78, 76 or 66 Gy, delivered with VMAT arcs of photons (6 or 10MV) on Novalis TX and Halcyon VARIAN®. We developed scripts to use MCO automatically and the AI algorithms were trained with 100 of our 78Gy treatment plans and they were adjusted in RayLearner to be used at 76 and 66Gy.

Plans were clinically accepted when they fullfil the RECORAD recommendations and pass the Patient Quality Assurance Criteria: Gamma index > 1 for 95% of points in 3% 2mm, measured with ARCHECK (and/or EPID SunCheck system (SunNuclear)

We compared MCO, ML and DL plans to manual Standard Optimized plans (SO) on 50 treatment plans with dosimetric Index: Conformity (C), Homogeneity (H) and Dose Gradient (DG). And the plan complexity were compared with the Modulation Complexity Score (MCS). We also evaluate the plan generation duration: active time (required by the user) and passive time (without the user). 

Results

SO plans were clinically accepted with index : C≈ 0.84, H≈0.07, DG ≈3.11, MSC ≈ 0.35 and plans were generated in an average time of 180 active minutes.

 

80% of ML plans were accepted after 16 passive minutes and 85% of DL plans were accepted after 11 passive minutes. 100 % of others AI plans were accepted after additionnal optimization (≈ 5 active minutes). AI dose distributions were at least equivalent and more complex  than SO: C≈ 0.86, H≈0.08, DG ≈3.02, MSC ≈ 0.31

 

100% of MCO plans were generated in 30 passive minutes and were after additionnal standard optimization (≈ 30 active minutes). MCO plans had the best and most complex dose distributions (C≈ 0.87, H≈0.06, DG ≈2.88, MSC ≈ 0.2)

Conclusion

Automated methods gives results more homogeneous and very close (DL / ML) or event better (MCO) than SO. Futhermore, calculation were 2(MCO) to 16 (DL) times faster. Finally, MCO seems to be more adapted to complex and unusual clinical cases and AI can be used on classical clinical cases which allows to consider adaptative radiotherapy.