A comprehensive daily QA routine at the ZAP-X system based on the RUBY phantom
MO-0053
Abstract
A comprehensive daily QA routine at the ZAP-X system based on the RUBY phantom
Authors: Katrin Buesing1, Joerg Harmsen2, Peter Douglas Klassen3, Hui Khee Looe4, Bjoern Poppe4, Daniela Poppinga5
1Carl-von-Ossietzky University , University Clinic for Medical Radiation Physics, Oldenburg, Germany; 2Practice for Radiation Therapy Nordhorn-Meppen, Radiation Therapy Meppen, Meppen, Germany; 3St. Bonifatius-Hospital , Department of Neurosurgery, Lingen, Germany; 4Carl-von-Ossietzky University, University Clinic for Medical Radiation Physics, Oldenburg, Germany; 5PTW Freiburg, PTW, Freiburg, Germany
Show Affiliations
Hide Affiliations
Purpose or Objective
The ZAP X
system is a novel system for stereotactic radiotherapy of brain lesions. It is
based on a gyroscopically mounted linear accelerator system with 3 MV nominal
photon energy with fixed collimators between 4 mm and 25 mm field size. The aim
of the work is the establishment of a fast and comprehensive daily check which
combines dosimetry and patient positioning in one workflow.
Material and Methods
In order to
establish the daily check, a planning CT of a RUBY head phantom (PTW Freiburg,
Germany) with RUBY System QA MultiMet insert (PTW Freiburg, Germany) and blind
plugs was performed once. For the planning CT, the standard patient
immobilization systems were used and a patient mask was prepared for the head
phantom. An isocenter was positioned at the central detector position and a treatment
plan was calculated using the 25 mm cone size and the “north pole” position
(equivalent to 0° at a conventional linear accelerator). The treatment plan
contains 500 MU comparable to the standard ZAP dosimetry routine in-air.
During the
daily routine, the RUBY system was positioned on the couch with the patient
mask and a Semiflex 3D ionization chamber (PTW Freiburg, Germany) was
positioned in the central detector position. The other two detector positions
were filled by blind plugs. The daily routine was started with the automated
patient positioning workflow. Thereby, the phantom is aligned by the dedicated
workflow based on kV images. After patient positioning, the treatment is
started and the dose is measured with the ionization chamber and a UNIDOS
electrometer (PTW Freiburg, Germany). The required couch displacement as well
as the measured dose is documented daily. As a comparison, the routine free in
air provided by ZAP was also performed daily with a Semiflex 3D chamber.
Results
Both methods,
dose measurement in air and the RUBY based workflow, showing constant dose
values within 1%. In contrast to the ZAP routine, the RUBY-based workflow can
also be used to check patient positioning daily. However, the temperature
correction for dose measurement must be observed for the RUBY phantom. For the
daily check, the RUBY phantom was stored in the treatment room and the
temperature in the treatment room was used as the correction temperature.
During the short time that the phantom is positioned in the ZAP system, the
phantom does not adapt to the temperature of the internal space of the ZAP system.
Conclusion
The RUBY
system can be used for a daily check on the ZAP system. It allows the
combination of dose check and patient positioning system check in one workflow.
This enables a fast and comprehensive check of the system in the morning.